Abstract: Purpose: Thermodilution continuous cardiac output measurements (TDCCO) by pulmonary artery catheter (PAC) have not been validated during therapeutic hypothermia in post-cardiac arrest patients. The calculated cardiac output based on the indirect Fick principle (FCO) using pulmonary artery blood gas mixed venous oxygen saturation (FCO-BG-SvO2) is considered as the gold standard. Continuous SvO2 by PAC (PAC-SvO2) has also not been validated previously during hypothermia. The aims of this study were (1) to compare FCO-BG-SvO2 with TDCCO, (2) to compare PAC-SvO2 with BG-SvO2 and finally (3) to compare FCO with SvO2 obtained via PAC or blood gas. Methods: We analyzed 102 paired TDCCO/FCO-BG-SvO2 and 88 paired BG-SvO2/PAC-SvO2 measurements in 32 post-cardiac arrest patients during therapeutic hypothermia. Results: TDCCO was significantly although poorly correlated with FCO-BG-SvO2 (R-2 0.21, p < 0.01) without systematic bias (-0.15 +/- 1.76 l/min). Analysis according to Bland and Altman however showed broad limits of agreement ([-3.61; 3.45] l/min) and an unacceptable high percentage error (105%). None of the criteria for clinical interchangeability were met. Concordance analysis showed that TDCCO had limited trending ability (R-2 0.03). FCO based on PAC-SvO2 was highly correlated with FCO-BG-SvO2 (R-2 0.72) with a small bias (-0.08 +/- 0.72 l/min) and slightly too high percentage error (44%). Conclusion: Our results show an extreme inaccuracy of TDCCO by PAC in post-cardiac arrest patients during therapeutic hypothermia. We found a reasonable correlation between BG-SvO2 and PAC-SvO2 and subsequently between FCO calculated with SvO2 obtained either via blood gas or PAC. The decision to start or titrate inotropics should therefore not be guided by TDCCO in this setting. (C) 2014 Elsevier Ireland Ltd. All rights reserved.
Notes: [Ameloot, K.; Meex, I.; Mullens, W.; Dens, J.; Dupont, M.] Ziekenhuis Oost Limburg, Dept Cardiol, B-3600 Genk, Belgium. [Meex, I.; Genbrugge, C.; Jans, F.; De Deyne, C.] Ziekenhuis Oost Limburg, Dept Anesthesiol & Crit Care Med, B-3600 Genk, Belgium. [Meex, I.; Genbrugge, C.; Jans, F.; Mullens, W.; Dens, J.; De Deyne, C.] Univ Hasselt, Fac Med & Life Sci, Diepenbeek, Belgium. [Malbrain, M.] ZNA Stuivenberg, Dept Intens Care Med, Antwerp, Belgium.
Keywords: continuous thermodilution cardiac output; post-cardiac arrest patients; therapeutic hypothermia;Continuous thermodilution cardiac output; Post-cardiac arrest patients; Therapeutic hypothermia
Document URI: http://hdl.handle.net/1942/17737
ISSN: 0300-9572
e-ISSN: 1873-1570
DOI: 10.1016/j.resuscitation.2014.06.025
ISI #: 000341306300045
Rights: © 2014 Elsevier Ireland Ltd. All rights reserved.
Category: A1
Type: Journal Contribution
Validations: ecoom 2015
Appears in Collections:Research publications

Files in This Item:
File Description SizeFormat 
ameloot 1.pdf
  Restricted Access
Published version781.87 kBAdobe PDFView/Open    Request a copy
Show full item record

SCOPUSTM   
Citations

8
checked on Sep 3, 2020

WEB OF SCIENCETM
Citations

14
checked on Apr 14, 2024

Page view(s)

64
checked on Jun 14, 2022

Download(s)

48
checked on Jun 14, 2022

Google ScholarTM

Check

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.