Please use this identifier to cite or link to this item: http://hdl.handle.net/1942/18710
Full metadata record
DC FieldValueLanguage
dc.contributor.authorBERTRAND, Philippe-
dc.contributor.authorGRIETEN, Lars-
dc.contributor.authorSMEETS, Christophe-
dc.contributor.authorVERBRUGGE, Frederik-
dc.contributor.authorMULLENS, Wilfried-
dc.contributor.authorVROLIX, Mathias-
dc.contributor.authorRivero-Ayerza, Maximo-
dc.contributor.authorVerhaert, David-
dc.contributor.authorVANDERVOORT, Pieter-
dc.date.accessioned2015-04-14T10:24:34Z-
dc.date.available2015-04-14T10:24:34Z-
dc.date.issued2015-
dc.identifier.citationECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES, 32 (3), p. 557-564-
dc.identifier.issn0742-2822-
dc.identifier.urihttp://hdl.handle.net/1942/18710-
dc.description.abstractPurpose: Echocardiography is increasingly important in the guidance and follow-up of percutaneous transcatheter device closures. It was recently shown that the Amplatzer left atrial appendage occluder frequently presents as a figure-of-eight artifact due to interaction of device mesh and ultrasound waves. It remains unknown whether this can be translated to other types of disc occluders. Furthermore, the morphology of this figure-of-eight artifact appears to be different in the transesophageal and transthoracic image of the same device. The aim of this study was to evaluate the echocardiographic appearance of different types of disc occluders, and to clarify differences in morphology of the figure-of-eight artifact. Methods: A mathematical model of an epitrochoid curve was used for numerical simulation of disc occluder appearance at various imaging depths. In addition, an in vitro setup was used for echocardiographic analysis of different types of disc occluders at adjustable imaging depth and position. Results: Mathematically, decreasing the imaging depth resulted in a more asymmetric figure-of-eight, i.e. with small upper part and wide lower part. In vitro results were in close agreement with the mathematical results. In addition, in vitro a figure-of-eight artifact was obtained in all different types of disc occluder devices. Conclusions: Different types of percutaneous disc occluders all present as a figure-of eight artifact on echocardiography when imaged from a coronal imaging position. The morphology of the artifact depends on the imaging depth, with a more asymmetric figure-of-eight morphology at smaller probe-to-device distance. This clarifies the differences observed between transesophageal and transthoracic imaging. (Echocardiography 2015;32:557–564)-
dc.description.sponsorshipP.B and F.V are supported by a grant of the Research Foundation - Flanders (FWO). This study is also part of the Limburg Clinical Research Program (LCRP) UHasselt-ZOL-Jessa, supported by the foundation Limburg Sterk Merk, Hasselt University, Ziekenhuis Oost-Limburg, and Jessa Hospital.-
dc.language.isoen-
dc.publisherWILEY-BLACKWELL-
dc.rights© 2014, Wiley Periodicals, Inc.-
dc.subject.otherultrasound physics; percutaneous closure devices; echocardiography; mathematical modeling; in vitro testing-
dc.subject.otherultrasound physics; percutaneous closure devices; echocardiography; mathematical modeling; in vitro testing-
dc.titleThe Figure-of-Eight Artifact in the Echocardiographic Assessment of Percutaneous Disc Occluders: Impact of Imaging Depth and Device Type-
dc.typeJournal Contribution-
dc.identifier.epage564-
dc.identifier.issue3-
dc.identifier.spage557-
dc.identifier.volume32-
local.format.pages8-
local.bibliographicCitation.jcatA1-
dc.description.notesVandervoort, PM (reprint author) [Bertrand, Philippe B.; Grieten, Lars; Verbrugge, Frederik H.; Mullens, Wilfried; Vrolix, Mathias; Rivero-Ayerza, Maximo; Verhaert, David; Vandervoort, Pieter M.] Ziekenhuis Oost Limburg, Dept Cardiol, B-3600 Genk, Belgium. [Bertrand, Philippe B.; Grieten, Lars; Smeets, Christophe J. P.; Verbrugge, Frederik H.; Mullens, Wilfried; Vandervoort, Pieter M.] Hasselt Univ, Fac Med & Life Sci, Hasselt, Belgium. pieter.vandervoort@zol.be-
local.publisher.placeHOBOKEN-
local.type.refereedRefereed-
local.type.specifiedArticle-
dc.identifier.doi10.1111/echo.12685-
dc.identifier.isi000350655900022-
item.fulltextWith Fulltext-
item.accessRightsRestricted Access-
item.fullcitationBERTRAND, Philippe; GRIETEN, Lars; SMEETS, Christophe; VERBRUGGE, Frederik; MULLENS, Wilfried; VROLIX, Mathias; Rivero-Ayerza, Maximo; Verhaert, David & VANDERVOORT, Pieter (2015) The Figure-of-Eight Artifact in the Echocardiographic Assessment of Percutaneous Disc Occluders: Impact of Imaging Depth and Device Type. In: ECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES, 32 (3), p. 557-564.-
item.validationecoom 2016-
item.contributorBERTRAND, Philippe-
item.contributorGRIETEN, Lars-
item.contributorSMEETS, Christophe-
item.contributorVERBRUGGE, Frederik-
item.contributorMULLENS, Wilfried-
item.contributorVROLIX, Mathias-
item.contributorRivero-Ayerza, Maximo-
item.contributorVerhaert, David-
item.contributorVANDERVOORT, Pieter-
crisitem.journal.issn0742-2822-
crisitem.journal.eissn1540-8175-
Appears in Collections:Research publications
Files in This Item:
File Description SizeFormat 
bertrand 1.pdf
  Restricted Access
Published version1.34 MBAdobe PDFView/Open    Request a copy
Show simple item record

Google ScholarTM

Check

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.