Please use this identifier to cite or link to this item:
http://hdl.handle.net/1942/19758
Title: | Accuracy of Dental Implant Placement Using CBCT-Derived Mucosa-Supported Stereolithographic Template | Authors: | SUN, Yi Luebbers, Heinz-Theo Agbaje, Jimoh Olubanwo Schepers, Serge POLITIS, Constantinus Van Slycke, Sarah Vrielinck, Luc |
Issue Date: | 2015 | Publisher: | WILEY-BLACKWELL | Source: | CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 17 (5), p. 862-870 | Abstract: | Purpose The aim of the present in vivo study was to evaluate whether a difference exists between the maxilla and the mandible regarding the precision of implant placement utilizing a cone beam computed tomography (CBCT)-derived mucosa-supported stereolithographic (SLA) template. Materials and MethodsEighty implants (44 maxilla, 36 mandible) were placed in 18 fully edentulous jaws (10 maxillas, eight mandibles) using a mucosa-supported SLA surgical template. A voxel-based registration technique was applied to match the postoperative and preoperative CBCT scans. ResultsVertical deviation (p=.026) at the implant hex and angular deviation (p=.0188) were significantly lower in the maxilla than in the mandible. The global linear deviation and lateral deviation at the implant hex were not significantly different. At the implant apex, the average maximum vertical deviation was within 1mm (0.1-4.6mm). The average maximum lateral deviation was 1.8mm (0.9-5.5mm) in the maxilla and 2.3mm (0.5-5.5mm) in the mandible when a 15-mm-long implant was placed. ConclusionsWhen using CBCT-derived mucosa-supported SLA templates, clinicians should be aware of differences in the angular deviation of the implants in the mandible and maxilla. The average maximum linear deviation should be considered as a safety margin at the implant apex. | Notes: | [Sun, Yi; Agbaje, Jimoh Olubanwo; Schepers, Serge; Van Slycke, Sarah; Vrielinck, Luc] Ziekenhuis Oost Limburg, Oral & Maxillofacial Surg, B-3600 Genk, Belgium. [Sun, Yi; Politis, Constantinus] Hasselt Univ, Fac Med, Diepenbeek, Belgium. [Luebbers, Heinz-Theo] Univ Zurich Hosp, Clin Craniomaxillofacial Surg, CH-8091 Zurich, Switzerland. [Agbaje, Jimoh Olubanwo; Politis, Constantinus] Katholieke Univ Leuven, Dept Oral Hlth Sci, Leuven, Belgium. [Luebbers, Heinz-Theo; Politis, Constantinus] Univ Hosp Leuven, Oral & Maxillofacial Surg, Leuven, Belgium. [Schepers, Serge] Univ Ghent, Fac Med, Oral & Maxillofacial Surg, B-9000 Ghent, Belgium. | Keywords: | accuracy; computer-assisted; cone beam computed tomography; dental implant; stereolithographic; surgical template;accuracy; computer-assisted; cone beam computed tomography; dental implant; stereolithographic; surgical template | Document URI: | http://hdl.handle.net/1942/19758 | ISSN: | 1523-0899 | e-ISSN: | 1708-8208 | DOI: | 10.1111/cid.12189 | ISI #: | 000362432900004 | Rights: | © 2013 Wiley Periodicals, Inc. | Category: | A1 | Type: | Journal Contribution | Validations: | ecoom 2016 |
Appears in Collections: | Research publications |
Files in This Item:
File | Description | Size | Format | |
---|---|---|---|---|
Sun_et_al-2015-Clinical_Implant_Dentistry_and_Related_Research.pdf Restricted Access | 470.57 kB | Adobe PDF | View/Open Request a copy |
SCOPUSTM
Citations
14
checked on Sep 2, 2020
WEB OF SCIENCETM
Citations
31
checked on Oct 14, 2024
Page view(s)
52
checked on Jun 14, 2022
Download(s)
36
checked on Jun 14, 2022
Google ScholarTM
Check
Altmetric
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.