Please use this identifier to cite or link to this item: http://hdl.handle.net/1942/21114
Full metadata record
DC FieldValueLanguage
dc.contributor.authorDESTEGHE, Lien-
dc.contributor.authorRaymaekers, Zina-
dc.contributor.authorLutin, Mark-
dc.contributor.authorVIJGEN, Johan-
dc.contributor.authorDilling-Boer, Dagmara-
dc.contributor.authorKOOPMAN, Pieter-
dc.contributor.authorSCHURMANS, Joris-
dc.contributor.authorVanduynhoven, Philippe-
dc.contributor.authorDENDALE, Paul-
dc.contributor.authorHEIDBUCHEL, Hein-
dc.date.accessioned2016-05-18T11:39:33Z-
dc.date.available2016-05-18T11:39:33Z-
dc.date.issued2016-
dc.identifier.citationEUROPACE, 19 (1), pag. 29-39-
dc.identifier.issn1099-5129-
dc.identifier.urihttp://hdl.handle.net/1942/21114-
dc.description.abstractAIMS: To determine the usability, accuracy, and cost-effectiveness of two handheld single-lead electrocardiogram (ECG) devices for atrial fibrillation (AF) screening in a hospital population with an increased risk for AF. METHODS AND RESULTS: Hospitalized patients (n = 445) at cardiological or geriatric wards were screened for AF by two handheld ECG devices (MyDiagnostick and AliveCor). The performance of the automated algorithm of each device was evaluated against a full 12-lead or 6-lead ECG recording. All ECGs and monitor tracings were also independently reviewed in a blinded fashion by two electrophysiologists. Time investments by nurses and physicians were tracked and used to estimate cost-effectiveness of different screening strategies. Handheld recordings were not possible in 7 and 21.4% of cardiology and geriatric patients, respectively, because they were not able to hold the devices properly. Even after the exclusion of patients with an implanted device, sensitivity and specificity of the automated algorithms were suboptimal (Cardiology: 81.8 and 94.2%, respectively, for MyDiagnostick; 54.5 and 97.5%, respectively, for AliveCor; Geriatrics: 89.5 and 95.7%, respectively, for MyDiagnostick; 78.9 and 97.9%, respectively, for AliveCor). A scenario based on automated AliveCor evaluation in patients without AF history and without an implanted device proved to be the most cost-effective method, with a provider cost to identify one new AF patient of €193 and €82 at cardiology and geriatrics, respectively. The cost to detect one preventable stroke per year would be €7535 and €1916, respectively (based on average CHA2DS2-VASc of 3.9 ± 2.0 and 5.0 ± 1.5, respectively). Manual interpretation increases sensitivity, but decreases specificity, doubling the cost per detected patient, but remains cheaper than sole 12-lead ECG screening. CONCLUSION: Using AliveCor or MyDiagnostick handheld recorders requires a structured screening strategy to be effective and cost-effective in a hospital setting. It must exclude patients with implanted devices and known AF, and requires targeted additional 12-lead ECGs to optimize specificity. Under these circumstances, the expenses per diagnosed new AF patient and preventable stroke are reasonable.-
dc.description.sponsorshipThis study is part of the Limburg Clinical Research Program (LCRP) UHasselt-ZOL-Jessa, supported by the foundation Limburg Sterk Merk, province of Limburg, Flemish government, Hasselt University, Ziekenhuis Oost-Limburg, and Jessa Hospital.-
dc.language.isoen-
dc.rights© The Author 2016. For permissions please email: journals.permissions@oup.com.-
dc.subject.otheratrial fibrillation; cost-effectiveness; electrocardiogram; handheld ECG; screening-
dc.titlePerformance of handheld electrocardiogram devices to detect atrial fibrillation in a cardiology and geriatric ward setting-
dc.typeJournal Contribution-
dc.identifier.epage39-
dc.identifier.issue1-
dc.identifier.spage29-
dc.identifier.volume19-
local.format.pages11-
local.bibliographicCitation.jcatA1-
dc.description.notesDesteghe, L (reprint author), Hasselt Univ, Fac Med & Life Sci, Martelarenlaan 42, B-3500 Hasselt, Belgium. lien.desteghe@uhasselt.be-
local.type.refereedRefereed-
local.type.specifiedArticle-
dc.identifier.doi10.1093/europace/euw025-
dc.identifier.isi000397055100005-
item.fulltextWith Fulltext-
item.accessRightsRestricted Access-
item.fullcitationDESTEGHE, Lien; Raymaekers, Zina; Lutin, Mark; VIJGEN, Johan; Dilling-Boer, Dagmara; KOOPMAN, Pieter; SCHURMANS, Joris; Vanduynhoven, Philippe; DENDALE, Paul & HEIDBUCHEL, Hein (2016) Performance of handheld electrocardiogram devices to detect atrial fibrillation in a cardiology and geriatric ward setting. In: EUROPACE, 19 (1), pag. 29-39.-
item.validationecoom 2018-
item.contributorDESTEGHE, Lien-
item.contributorRaymaekers, Zina-
item.contributorLutin, Mark-
item.contributorVIJGEN, Johan-
item.contributorDilling-Boer, Dagmara-
item.contributorKOOPMAN, Pieter-
item.contributorSCHURMANS, Joris-
item.contributorVanduynhoven, Philippe-
item.contributorDENDALE, Paul-
item.contributorHEIDBUCHEL, Hein-
crisitem.journal.issn1099-5129-
crisitem.journal.eissn1532-2092-
Appears in Collections:Research publications
Files in This Item:
File Description SizeFormat 
desteghe 1.pdf
  Restricted Access
Published version642.36 kBAdobe PDFView/Open    Request a copy
Show simple item record

SCOPUSTM   
Citations

64
checked on Sep 3, 2020

WEB OF SCIENCETM
Citations

87
checked on May 9, 2024

Page view(s)

66
checked on Aug 10, 2022

Download(s)

46
checked on Aug 10, 2022

Google ScholarTM

Check

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.