Please use this identifier to cite or link to this item: http://hdl.handle.net/1942/23094
Title: The first clinical experience with a novel "locking" microcatheter in chronic coronary total occlusions
Authors: Wilson, Simon J.
MAEREMANS, Joren 
Walsh, Simon J.
Hanratty, Colm G.
Meyer-Gessner, Markus
DENS, Jo 
Christiansen, Evald H.
Bufe, Alexander
Spratt, James C.
Issue Date: 2017
Source: EuroIntervention, 12(15), p.1883-1888
Abstract: A novel "locking" microcatheter has been developed to address residual failure modes in chronic total occlusion (CTO) percutaneous coronary intervention (PCI). The aim of this study was to report the first clinical experience of this device. METHODS AND RESULTS: The microcatheter was assessed prospectively in 92 unselected CTO cases across six European sites. Overall technical success was 85.9% (79/92) and 97.5% for patients with a J-CTO score of <3 (39/40). Within-CTO tortuosity was the only lesion characteristic to predict reduced technical success (OR 0.10 [0.01-0.97], p=0.047). Calcification (OR 0.45 [0.04-5.31], p=0.53), lesion length >20 mm (OR 0.58 [0.05-6.81], p=0.66) and a blunt proximal cap (OR 0.47 [0.08-2.90], p=0.42) were not associated with technical failure in this case series. Locking facilitated guidewire crossing (after initial failure) of the proximal cap in 23 cases, distal cap in 11 cases and CTO body in 10 cases, and delivery of the microcatheter through to the distal vessel in 22 cases. The overall procedural complication rate was 1.1% (1/92) with no major events. CONCLUSIONS: Guidewire locking with this novel microcatheter assists penetration and crossing of fibrocalcific anatomy with a high degree of safety. Using this device in CTO PCI may lead to improved primary wiring and overall procedural success rates.
Notes: Wilson, SJ (reprint author), Room SU305,Chancellors Bldg,49 Little France Cres, Edinburgh EH16 4SB, Midlothian, Scotland. simonwilson3@nhs.net
Keywords: stable angina; chronic coronary total occlusion; other technique
Document URI: http://hdl.handle.net/1942/23094
ISSN: 1774-024X
e-ISSN: 1969-6213
DOI: 10.4244/EIJ-D-16-00272
ISI #: 000397597600017
Rights: © Europa Digital & Publishing 2017. All rights reserved.
Category: A1
Type: Journal Contribution
Validations: ecoom 2018
Appears in Collections:Research publications

Files in This Item:
File Description SizeFormat 
The First Clinical Experience with a Novel Locking Microcatheter in Chronic Coronary Total Occlusions (1).pdf
  Restricted Access
Published version752.85 kBAdobe PDFView/Open    Request a copy
NHancer-EuroIntervention-Submitted paper.pdf
  Restricted Access
Peer-reviewed author version658.7 kBAdobe PDFView/Open    Request a copy
Show full item record

SCOPUSTM   
Citations

1
checked on Sep 3, 2020

WEB OF SCIENCETM
Citations

2
checked on Apr 22, 2024

Page view(s)

54
checked on Sep 7, 2022

Download(s)

76
checked on Sep 7, 2022

Google ScholarTM

Check

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.