dc.description.abstractBackground: During implantation of an uncemented femoral knee implant, press-fit interference fit provides the primary stability. It is assumed that during implantation a combination of elastic and plastic deformation and abrasion of the bone will occur, but little is known about what happens at the bone-implant interface and how much press-fit interference fit is eventually achieved. Methods: Five cadaveric femora were prepared and implantation was performed by an experienced surgeon. Micro-CT- and conventional CT-scans were obtained pre- and post-implantation for geometrical measurements and to measure bone mineral density. Additionally, the position of the implant with respect to the bone was determined by optical scanning of the reconstructions. By measuring the differences in surface geometry, assessments were made of the cutting error, the actual interference fit, the amount of bone damage, and the effective interference fit. Findings: Our analysis showed an average cutting error of 0.67 mm (SD 0.17 mm), which pointed mostly towards bone under-resections. We found an average actual AP interference fit of 1.48 mm (SD 0.27 mm), which was close to the nominal value of 1.5 mm. Interpretation: We observed combinations of bone damage and elastic deformation in all bone specimens, which showed a trend to be related with bone density. Higher bone density tended to lead to lower bone damage and higher elastic deformation. The results of the current study indicate different factors that interact while implanting an uncemented femoral knee component. This knowledge can be used to fine-tune design criteria of femoral components to achieve adequate primary stability for all patients.-
dc.description.sponsorshipThis work was supported by a research grant from DePuy Synthes Joint Reconstruction (Leeds, UK). The authors would like to thank orthopaedic surgeon, Dr. S. van de Groes, for performing implantation and Mr. F. Baan for his help with the optical scanning.-
dc.language.isoen-
dc.publisherELSEVIER SCI LTD-
dc.subject.otherUncemented total knee replacement; Bone damage; Micro-CT scan; Bone density; Press-fit interference fit-
dc.subject.otherUncemented total knee replacement; Bone damage; Micro-CT scan; Bone density; Press-fit interference fit-
dc.titleEvaluation of interference fit and bone damage of an uncemented femoral knee implant-
dc.typeJournal Contribution-
dc.identifier.epage9-
dc.identifier.spage1-
dc.identifier.volume51-
local.format.pages9-
local.bibliographicCitation.jcatA1-
dc.description.notes[Berahmani, Sanaz; Hendriks, Maartje; Janssen, Dennis; Verdonschot, Nico] Radboud Univ Nijmegen, Med Ctr, Radboud Inst Hlth Sci, Orthopaed Res Lab, POB 9101, NL-6500 HB Nijmegen, Netherlands. [de Jong, Joost J. A.; van den Bergh, Joop P. W.] Maastricht Univ, Med Ctr, NUTRIM Sch Nutr & Translat Res Metab, Maastricht, Netherlands. [de Jong, Joost J. A.; van den Bergh, Joop P. W.] Maastricht Univ, Med Ctr, Dept Rheumatol, Maastricht, Netherlands. [van den Bergh, Joop P. W.] VieCuri Med Ctr, Dept Internal Med, Venlo, Netherlands. [van den Bergh, Joop P. W.] Hasselt Univ, Fac Med & Life Sci, Hasselt, Belgium. [Maal, Thomas] Radboud Univ Nijmegen, Med Ctr, Dept Oral & Maxillofacial Surg, Nijmegen, Netherlands. [Verdonschot, Nico] Univ Twente, Lab Biomech Engn, Enschede, Netherlands.-
local.publisher.placeOXFORD-
local.type.refereedRefereed-
local.type.specifiedArticle-
dc.identifier.doi10.1016/j.clinbiomech.2017.10.022-
dc.identifier.isi000426235700001-
item.fulltextWith Fulltext-
item.fullcitationBerahmani, Sanaz; Hendriks, Maartje; de Jong, Joost J. A.; VAN DEN BERGH, Joop; Maal, Thomas; Janssen, Dennis & Verdonschot, Nico (2018) Evaluation of interference fit and bone damage of an uncemented femoral knee implant. In: CLINICAL BIOMECHANICS, 51, p. 1-9.-
item.contributorBerahmani, Sanaz-
item.contributorHendriks, Maartje-
item.contributorde Jong, Joost J. A.-
item.contributorVAN DEN BERGH, Joop-
item.contributorMaal, Thomas-
item.contributorJanssen, Dennis-
item.contributorVerdonschot, Nico-
item.accessRightsRestricted Access-
item.validationecoom 2019-
crisitem.journal.issn0268-0033-
crisitem.journal.eissn1879-1271-
Appears in Collections:Research publications
Files in This Item:
File Description SizeFormat 
berahmani 1.pdf
  Restricted Access
Published version2.14 MBAdobe PDFView/Open    Request a copy
Show simple item record

SCOPUSTM   
Citations

3
checked on Sep 2, 2020

WEB OF SCIENCETM
Citations

10
checked on May 10, 2024

Page view(s)

76
checked on Sep 7, 2022

Download(s)

42
checked on Sep 7, 2022

Google ScholarTM

Check

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.