Please use this identifier to cite or link to this item: http://hdl.handle.net/1942/36300
Full metadata record
DC FieldValueLanguage
dc.contributor.authorCottens, D-
dc.contributor.authorMAEREMANS, Joren-
dc.contributor.authorVROLIX, Mathias-
dc.contributor.authorVAN LIERDE, Johan-
dc.contributor.authorDENS, Jo-
dc.contributor.authorFerdinande, B-
dc.date.accessioned2021-12-20T14:38:22Z-
dc.date.available2021-12-20T14:38:22Z-
dc.date.issued2022-
dc.date.submitted2021-08-25T13:08:27Z-
dc.identifier.citationACTA CARDIOLOGICA, 77 (4) , p. 322-327-
dc.identifier.urihttp://hdl.handle.net/1942/36300-
dc.description.abstractBackground This study aimed to assess the reliability of pressure derived fractional flow reserve (FFR) measurement and the handling performance of the OptoWire Deux with an optical pressure sensor relative to both the PressureWire X and the Verrata Pressure wire with piezoresistive pressure sensors. Methods This single centre study included 80 patients between October 2016 and May 2017 undergoing a diagnostic coronary angiography. The reliability of FFR measurement measured with the OptoWire Deux relative to the PressureWire X and Verrata Pressure wire was assessed by the presence of drift. Drift was defined as a Pd/Pa measurement different from 1.00 +/- 0.02 when pulled back after a FFR measurement at the location of the initial equalisation. Handling characteristics for all pressure wires were assessed qualitatively with respect to the PressureWire Aeris. Results Ninety-eight measurements in 78 patients were performed; two patients were excluded because the lesion could not be crossed. Very stable and reliable FFR measurements with the optical sensors were registered, relative to the piezoresistive pressure sensors. Drift was found in 11%, 37%, and 33% of the measurements for OptoWire Deux, PressureWire X, and Verrata Pressure wire respectively. The handling performance of the OptoWire Deux was better rated for steerability and torqueability in relation to the standard FFR wire. The handling of the PressureWire X was rated equally good whereas the handling of the Verrata pressure wire was rated inferior in relation to the standard FFR wire. Conclusions In patients undergoing FFR measurement, the OptoWire Deux has a stable and reliable pressure hence FFR measurement with fewer drift events and has good handling characteristics.-
dc.description.sponsorshipThe authors thank Marcel van ‘t Veer from the Department of Biomedical Engineering of the University of Technology Eindhoven for his statistical support. Jo Dens receives grants from TopMedical (Distributor of OptoWire Deux)Asahi Intecc co. materials) for teaching courses and proctoring. Joren Maeremans is researcher for the Limburg Clinical Research Program UHasselt-ZOL-Jessa, supported by the foundation Limburg Sterk Merk, Hasselt University, Ziekenhuis Oost-Limburg and Jessa Hospital.-
dc.language.isoen-
dc.publisherTAYLOR & FRANCIS LTD-
dc.subject.otherFractional flow reserve-
dc.subject.otherdrift-
dc.subject.otheroptical pressure sensor-
dc.subject.otherpiezoresistive pressure sensor-
dc.titleFFR pressure wire comparative study: piezoresistive versus optical sensor-
dc.typeJournal Contribution-
dc.identifier.epage327-
dc.identifier.issue4-
dc.identifier.spage322-
dc.identifier.volume77-
local.bibliographicCitation.jcatA1-
local.publisher.place2-4 PARK SQUARE, MILTON PARK, ABINGDON OR14 4RN, OXON, ENGLAND-
local.type.refereedRefereed-
local.type.specifiedArticle-
dc.identifier.doi10.1080/00015385.2021.1939510-
dc.identifier.pmid34218723-
dc.identifier.isi000669782600001-
local.provider.typeWeb of Science-
local.uhasselt.internationalno-
item.validationecoom 2022-
item.fulltextNo Fulltext-
item.accessRightsClosed Access-
item.fullcitationCottens, D; MAEREMANS, Joren; VROLIX, Mathias; VAN LIERDE, Johan; DENS, Jo & Ferdinande, B (2022) FFR pressure wire comparative study: piezoresistive versus optical sensor. In: ACTA CARDIOLOGICA, 77 (4) , p. 322-327.-
item.contributorCottens, D-
item.contributorMAEREMANS, Joren-
item.contributorVROLIX, Mathias-
item.contributorVAN LIERDE, Johan-
item.contributorDENS, Jo-
item.contributorFerdinande, B-
crisitem.journal.issn0001-5385-
crisitem.journal.eissn1784-973X-
Appears in Collections:Research publications
Show simple item record

WEB OF SCIENCETM
Citations

1
checked on Apr 30, 2024

Page view(s)

44
checked on Aug 10, 2022

Google ScholarTM

Check

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.