Please use this identifier to cite or link to this item: http://hdl.handle.net/1942/37165
Full metadata record
DC FieldValueLanguage
dc.contributor.authorCottens, Daan-
dc.contributor.authorFerdinande, Bert-
dc.contributor.authorPolad, Jawed-
dc.contributor.authorVROLIX, Mathias-
dc.contributor.authorAMELOOT, Koen-
dc.contributor.authorHendrickx, Ief-
dc.contributor.authorPoels , Ella-
dc.contributor.authorMAEREMANS, Joren-
dc.contributor.authorDENS, Jo-
dc.date.accessioned2022-04-04T13:22:33Z-
dc.date.available2022-04-04T13:22:33Z-
dc.date.issued2022-
dc.date.submitted2022-03-31T11:05:21Z-
dc.identifier.citationAmerican Journal of Cardiovascular Disease, 12 (1) , p. 42 -52-
dc.identifier.urihttp://hdl.handle.net/1942/37165-
dc.description.abstractBackground: This study aimed to assess the stability of pressure derived fractional flow reserve (FFR) measurement and the handling performance of the OptoWire Deux with an optical pressure sensor relative to the PressureWire X with piezo resistive pressure sensors. Methods: This multicenter centre observational study included 50 patients between June 2017 and November 2018 undergoing a diagnostic coronary angiography with FFR measurement of moderate to severe lesions. The reliability of FFR measurement measured with the OptoWire Deux relative to the PressureWire X in each lesion was assessed by the presence of drift. Handling characteristics for both pressure wires were assessed by a 5-point scale and by comparing the time between equalization and crossing the distal target lesion. Results: Hundred and sixteen measurements in 50 patients were performed. Very stable and reliable FFR measurements with the optical sensors were registered, relative to the piezo resistive pressure sensors. There is statistically significant difference in favor of the OptoWire Deux over the PressureWire X (P=0.001). However, the differences are small, when drift values were compared as continuous variables, no statistically significant difference was found for both directional (P=0.435) as for absolute drift (P=0.058). Conclusions: In patients undergoing FFR measurement, both optical sensor pressure wires (Optowire Deux) as piezo resistive sensor pressure wires (PressureWire X) generate stable and reliable pressure and thus FFR measurement. The optical pressure sensor is less susceptible for drift relative to the piezo resistive pressure sensor, but the difference is within an acceptable range.-
dc.description.sponsorshipWe would like to thank Jo Zelis and Marcel van ‘t Veer from the Department of Biomedical Engineering of the University of Technology Eindhoven for his statistical support.-
dc.language.isoen-
dc.publisherE-CENTURY PUBLISHING CORP-
dc.subject.otherFractional flow reserve-
dc.subject.otherdrift-
dc.subject.otheroptical pressure sensor-
dc.subject.otherpiezo resistive pressure sensor-
dc.titleFFR pressure wire comparative study for drift: piezo resistive versus optical sensor-
dc.typeJournal Contribution-
dc.identifier.epage52-
dc.identifier.issue1-
dc.identifier.spage42-
dc.identifier.volume12-
local.format.pages11-
local.bibliographicCitation.jcatA1-
dc.description.notesDens, J (corresponding author), Hosp Oost Limburg, Dept Cardiol, Schiepse Bos 6, Genk, Belgium.-
dc.description.notesjo.dens@zol.be-
local.publisher.place40 WHITE OAKS LN, MADISON, WI 53711 USA-
local.type.refereedRefereed-
local.type.specifiedArticle-
dc.identifier.isiWOS:000767769000005-
local.provider.typewosris-
local.description.affiliation[Cottens, Daan; Ferdinande, Bert; Vrolix, Mathias; Ameloot, Koen; Hendrickx, Ief; Poels, Ella; Dens, Jo] Hosp Oost Limburg, Dept Cardiol, Schiepse Bos 6, Genk, Belgium.-
local.description.affiliation[Maeremans, Joren; Dens, Jo] Univ Hasselt, Fac Med & Life Sci, Hasselt, Belgium.-
local.description.affiliation[Polad, Jawed] Hosp Jeroen Bosch, Dept Cardiol, Shertogenbosch, Netherlands.-
local.uhasselt.internationalyes-
item.contributorCottens, Daan-
item.contributorFerdinande, Bert-
item.contributorPolad, Jawed-
item.contributorVROLIX, Mathias-
item.contributorAMELOOT, Koen-
item.contributorHendrickx, Ief-
item.contributorPoels , Ella-
item.contributorMAEREMANS, Joren-
item.contributorDENS, Jo-
item.fullcitationCottens, Daan; Ferdinande, Bert; Polad, Jawed; VROLIX, Mathias; AMELOOT, Koen; Hendrickx, Ief; Poels , Ella; MAEREMANS, Joren & DENS, Jo (2022) FFR pressure wire comparative study for drift: piezo resistive versus optical sensor. In: American Journal of Cardiovascular Disease, 12 (1) , p. 42 -52.-
item.accessRightsRestricted Access-
item.fulltextWith Fulltext-
crisitem.journal.issn2160-200X-
Appears in Collections:Research publications
Files in This Item:
File Description SizeFormat 
ajcd0012-0042.pdf
  Restricted Access
Published version646.29 kBAdobe PDFView/Open    Request a copy
Show simple item record

WEB OF SCIENCETM
Citations

2
checked on May 2, 2024

Page view(s)

60
checked on Sep 6, 2022

Google ScholarTM

Check


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.