Please use this identifier to cite or link to this item: http://hdl.handle.net/1942/47563
Full metadata record
DC FieldValueLanguage
dc.contributor.authorBonne, Lawrence-
dc.contributor.authorBergans, Niki-
dc.contributor.authorMccutcheon, Keir-
dc.contributor.authorLAENEN, Annouschka-
dc.contributor.authorBogaerts, Ria-
dc.contributor.authorVanheule, Jeroen-
dc.contributor.authorBennett, Johan-
dc.contributor.authorCockmartin, Lesley-
dc.contributor.authorBosmans, Hilde-
dc.contributor.authorDesmet, Walter-
dc.contributor.authorMaleux, Geert-
dc.date.accessioned2025-10-20T13:04:44Z-
dc.date.available2025-10-20T13:04:44Z-
dc.date.issued2025-
dc.date.submitted2025-10-17T14:52:41Z-
dc.identifier.citationCardiovascular and interventional radiology,-
dc.identifier.urihttp://hdl.handle.net/1942/47563-
dc.description.abstractPurposeInterventional radiologists are among the healthcare professionals most exposed to ionizing scatter radiation. A 0.5 mm lead equivalent X-ray protective drape (MXPD, Mavig, Germany) was introduced to reduce operator radiation dose and has demonstrated efficacy during cardiac interventions. This study evaluates the efficacy of the MXPD during transfemoral liver interventions.Materials & MethodsA single-center randomized controlled study was conducted to compare operator radiation exposure during transfemoral liver interventions (n = 100) with and without the MXPD. Primary outcome was the ratio of the combined operator radiation dose values-measured at the chest, eyes and both hands and normalized to the respective dose area products - between MXPD and control groups. Secondary outcomes included the ratios in operator radiation doses measured for the individual dosimeter locations.ResultsThe MXPD group showed a significantly lower normalized radiation dose per DAP (mean reduction 0.31 mu Sv; 95% CI 0.083-0.536, p = 0.009), representing a 43% reduction. The effect of the drape was the highest for the whole-body dose measured above the lead apron (55% reduction, p = 0.10), the left hand finger dose (43%, p = 0.20) and the eye lens dose estimated at the left temple (63%, p = 0.10) and in between the eyes (47%, p = 0.10).ConclusionThe MXPD can effectively reduce operator radiation dose during transfemoral arterial liver interventions, offering a practical adjunct to standard protective measures.Level of EvidenceLevel 1b, Randomized Controlled Trial.-
dc.language.isoen-
dc.publisherSPRINGER-
dc.rightsSpringer Science+Business Media, LLC, part of Springer Nature and the Cardiovascular and Interventional Radiological Society of Europe (CIRSE) 2025-
dc.subject.otherTransarterial liver interventions-
dc.subject.otherRadiation protection-
dc.subject.otherPelvic drapes-
dc.titleEfficacy of an X-Ray Protective Drape in Reducing Operator Radiation Dose in Transfemoral Arterial Liver Interventions: a Randomized-Controlled Trial-
dc.typeJournal Contribution-
local.format.pages7-
local.bibliographicCitation.jcatA1-
dc.description.notesBonne, L (corresponding author), Univ Hosp Leuven, Dept Radiol, Herestr 49, B-3000 Louvain, Belgium.-
dc.description.noteslawrence.bonne@uzleuven.be-
local.publisher.placeONE NEW YORK PLAZA, SUITE 4600, NEW YORK, NY, UNITED STATES-
local.type.refereedRefereed-
local.type.specifiedArticle-
local.bibliographicCitation.statusEarly view-
dc.identifier.doi10.1007/s00270-025-04223-3-
dc.identifier.pmid41047378-
dc.identifier.isiWOS:001586551000001-
local.provider.typewosris-
local.description.affiliation[Bonne, Lawrence; Cockmartin, Lesley; Bosmans, Hilde; Maleux, Geert] Univ Hosp Leuven, Dept Radiol, Herestr 49, B-3000 Louvain, Belgium.-
local.description.affiliation[Bergans, Niki] Univ Hosp Leuven, Dept Oncol, Louvain, Belgium.-
local.description.affiliation[Bergans, Niki; Bogaerts, Ria; Vanheule, Jeroen] Univ Hosp Leuven, Radiat Protect & Dosimetry Serv, Dept Hlth Safety & Environm, Leuven, Belgium.-
local.description.affiliation[Mccutcheon, Keir; Bennett, Johan; Desmet, Walter] Univ Hosp Leuven, Dept Cardiovasc Med, Louvain, Belgium.-
local.description.affiliation[Laenen, Annouschka] Interuniv Inst Biostat & Stat Bioinformat, Louvain, Belgium.-
local.description.affiliation[Bosmans, Hilde] Katholieke Univ Leuven, Dept Imaging & Pathol, Div Med Phys & Qual Assessment, Louvain, Belgium.-
local.uhasselt.internationalno-
item.accessRightsRestricted Access-
item.fulltextWith Fulltext-
item.contributorBonne, Lawrence-
item.contributorBergans, Niki-
item.contributorMccutcheon, Keir-
item.contributorLAENEN, Annouschka-
item.contributorBogaerts, Ria-
item.contributorVanheule, Jeroen-
item.contributorBennett, Johan-
item.contributorCockmartin, Lesley-
item.contributorBosmans, Hilde-
item.contributorDesmet, Walter-
item.contributorMaleux, Geert-
item.fullcitationBonne, Lawrence; Bergans, Niki; Mccutcheon, Keir; LAENEN, Annouschka; Bogaerts, Ria; Vanheule, Jeroen; Bennett, Johan; Cockmartin, Lesley; Bosmans, Hilde; Desmet, Walter & Maleux, Geert (2025) Efficacy of an X-Ray Protective Drape in Reducing Operator Radiation Dose in Transfemoral Arterial Liver Interventions: a Randomized-Controlled Trial. In: Cardiovascular and interventional radiology,.-
crisitem.journal.issn0174-1551-
crisitem.journal.eissn1432-086X-
Appears in Collections:Research publications
Files in This Item:
File Description SizeFormat 
Efficacy of an X-Ray Protective Drape in Reducing Operator .pdf
  Restricted Access
Early view1.16 MBAdobe PDFView/Open    Request a copy
Show simple item record

Google ScholarTM

Check

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.