Please use this identifier to cite or link to this item:
http://hdl.handle.net/1942/47712| Title: | Polyethylene liner fracture in dual-mobility trapeziometacarpal joint prosthesis: Report of 17 cases | Authors: | DUERINCKX, Joris Marks, Miriam Herren, Daniel B. |
Issue Date: | 2025 | Publisher: | SAGE PUBLICATIONS LTD | Source: | Journal of hand surgery. European volume, | Status: | Early view | Abstract: | We report 17 polyethylene liner fractures in 15 patients with dual mobility trapeziometacarpal joint prostheses who were primarily active males with high key pinch strength. Treatment involved head and liner exchange or trapeziectomy, depending on implant damage severity.Evidence level: V | Notes: | Duerinckx, J (corresponding author), Ziekenhuis Oost Limburg, Dept Orthopaed Surg & Traumatol, Synaps Pk 1, B-3600 Genk, Belgium. joris.duerinckx@zol.be |
Keywords: | complication;Moovis;revision surgery;Touch;trapeziometacarpal osteoarthritis | Document URI: | http://hdl.handle.net/1942/47712 | ISSN: | 1753-1934 | e-ISSN: | 2043-6289 | DOI: | 10.1177/17531934251383007 | ISI #: | 001599637800001 | Rights: | The Author(s) 2025 | Category: | A1 | Type: | Journal Contribution |
| Appears in Collections: | Research publications |