dc.description.abstractBackground and purpose: Remote beam output audits (BOAs) are important for verifying the calibration of the audited beam. However, BOAs (Level I audit) provide limited independent validation of the commissioned beam model in the treatment planning system (TPS). With Level II audits, the beam commissioning data is checked more thoroughly. This study presents the results of a remote audit program of reference and non-reference dosimetry, utilising alanine dosimetry. A particularity was to use the calculations of the TPS as reference values corrected by the daily output. The goal was to verify the beam commissioning data in the TPS, including the calibration. Materials and methods: The audit consisted of five irradiations with a 10 x 10 cm2 field and a small field test with field sizes up to 2 x 2 cm2. Watertight sealed alanine detectors, along with a 3D-printed holder featuring three slots at depths of 8 cm, 10 cm, and 20 cm, were used. The results of 224 beams audited were included in this study. Results: The difference between the measured and predicted doses was within +/- 5 % for all beams except one. No clear trend was observed in the results for uniform and FFF beams. The uncertainty in the alanine readout was 1 % (k =1). The results maintained their accuracy, while a slight improvement was observed in the small field test. Conclusions: All audit results were satisfactory. There is a slightly positive trend in the audit results over the years, suggesting that the accuracy of radiotherapy calculations is consistently maintained.-
dc.description.sponsorshipThe BELdART project is sponsored by Het Kankerplan (the Cancer plan) and is monitored by the Belgian College of Radiation Oncology between 2009-2025. A steering committee consisting of five senior medical physics experts provide guidance to the BELdART project.-
dc.language.isoen-
dc.publisherELSEVIER-
dc.rights2026 The Authors. Published by Elsevier B.V. on behalf of European Society of Radiotherapy & Oncology. This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).-
dc.subject.otherAlanine/EPR dosimetry-
dc.subject.otherDosimetry audit-
dc.subject.otherQA-
dc.subject.otherBeam output-
dc.titleResults of a remote audit program of reference and non-reference dosimetry of photon beams using alanine detectors-
dc.typeJournal Contribution-
dc.identifier.volume37-
local.format.pages7-
local.bibliographicCitation.jcatA1-
dc.description.notesYalvac, B (corresponding author), Univ Hasselt, CMK, NuTeC, Diepenbeek, Belgium.-
dc.description.notesburak.yalvac@uhasselt.be; brigitte.reniers@uhasselt.be-
local.publisher.placeRADARWEG 29, 1043 NX AMSTERDAM, NETHERLANDS-
local.type.refereedRefereed-
local.type.specifiedArticle-
local.bibliographicCitation.artnr100914-
dc.identifier.doi10.1016/j.phro.2026.100914-
dc.identifier.pmid41704950-
dc.identifier.isi001686467600001-
local.provider.typewosris-
local.description.affiliation[Yalvac, Burak; Reulens, Nathalie; Reniers, Brigitte] Univ Hasselt, CMK, NuTeC, Diepenbeek, Belgium.-
local.uhasselt.internationalno-
item.fulltextWith Fulltext-
item.contributorYALVAC, Burak-
item.contributorREULENS, Nathalie-
item.contributorRENIERS, Brigitte-
item.fullcitationYALVAC, Burak; REULENS, Nathalie & RENIERS, Brigitte (2026) Results of a remote audit program of reference and non-reference dosimetry of photon beams using alanine detectors. In: Physics & Imaging in Radiation Oncology, 37 (Art N° 100914).-
item.accessRightsOpen Access-
crisitem.journal.eissn2405-6316-
Appears in Collections:Research publications
Files in This Item:
File Description SizeFormat 
main.pdfPublished version3.91 MBAdobe PDFView/Open
Show simple item record

Google ScholarTM

Check

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.