Title: | Image quality vs radiation dose of four cone beam computed tomography scanners | Authors: | Loubele, Miet Jacobs, Reinhilde Maes, Frederik Denis, Kathleen White, Stewart Coudyzer, W LAMBRICHTS, Ivo van Steenberghe, Daniel Seutens, Paul |
Issue Date: | 2008 | Publisher: | Britisch Institute of Radiology | Source: | DENTOMAXILLOFACIAL RADIOLOGY, 37. p. 309-318 | Abstract: | Objectives: To evaluate image quality by examining segmentation accuracy and assess radiation dose for cone beam CT (CBCT) scanners. Methods: A skull phantom, scanned by a laser scanner, and a contrast phantomwere used to evaluate segmentation accuracy. The contrast phantom consisted of a polymethyl methacrylate (PMMA) cylinder with cylindrical inserts of air, bone and PMMA. The phantoms were scanned on the (1) Accuitomo 3D, (2) MercuRay, (3) NewTom 3G, (4) i-CAT and (5) Sensation 16. The structures were segmented with an optimal treshold. Thicknesses of the bone of the mandible and the diameter of the cylinders in the contrast phantom were measured across lines at corresponding places in the contrast phantom were measured across lines at corresponding places in the CT image vs a ground truth. The accuracy was in the 95th percentile of the difference between corresponding measurements. The correlation between accuracy in skull and contrast phantom was calculated. The radiation dose was assessed by DPI100,c) (dose profile integral 100,c) at the central hole of a CT dose index (CTDI) phantom. Results: The results for the DPI100,c were 107 mGy mm for (1), 1569 mGy mm for (2), 446 mGy mm for (3), 249 mGy mm for (4) and 1090 mGy mm for (5). The segmentations in the contrast phantom were submillimeter accurate in all scanners. The segmentation accuracy of the mandible was 2.9 mm for (1), 4.2 mm for (2), 3.4 mm for (3), 1.0 mm for (4) and 1.2 mm for (5). The correlation between measurements in the contrast and skull phantom was below 0.37 mm. Conclusions: The best radiation dose vs image quality was found for the i-CAT. | Document URI: | http://hdl.handle.net/1942/8501 | ISSN: | 0250-832X | e-ISSN: | 1476-542X | DOI: | 10.1259/dmfr/16770531 | ISI #: | 000259375000002 | Category: | A1 | Type: | Journal Contribution | Validations: | ecoom 2009 |
Appears in Collections: | Research publications |
Show full item record
SCOPUSTM
Citations
90
checked on Sep 2, 2020
WEB OF SCIENCETM
Citations
98
checked on Oct 12, 2024
Page view(s)
96
checked on Jul 28, 2023
Google ScholarTM
Check
Altmetric
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.