Please use this identifier to cite or link to this item: http://hdl.handle.net/1942/17753
Title: The use of tetrahedral mesh geometries in Monte Carlo simulation of applicator based brachytherapy dose distributions
Authors: Fonseca, Gabriel Paiva
Landry, Guillaume
White, Shane
D'Amours, Michel
Yoriyaz, Helio
Beaulieu, Luc
RENIERS, Brigitte 
Verhaegen, Frank
Issue Date: 2014
Publisher: IOP PUBLISHING LTD
Source: PHYSICS IN MEDICINE AND BIOLOGY, 59 (19), p. 5921-5935
Abstract: Accounting for brachytherapy applicator attenuation is part of the recommendations from the recent report of AAPM Task Group 186. To do so, model based dose calculation algorithms require accurate modelling of the applicator geometry. This can be non-trivial in the case of irregularly shaped applicators such as the Fletcher Williamson gynaecological applicator or balloon applicators with possibly irregular shapes employed in accelerated partial breast irradiation (APBI) performed using electronic brachytherapy sources (EBS). While many of these applicators can be modelled using constructive solid geometry (CSG), the latter may be difficult and time-consuming. Alternatively, these complex geometries can be modelled using tessellated geometries such as tetrahedral meshes (mesh geometries (MG)). Recent versions of Monte Carlo (MC) codes Geant4 and MCNP6 allow for the use of MG. The goal of this work was to model a series of applicators relevant to brachytherapy using MG. Applicators designed for Ir-192 sources and 50 kV EBS were studied; a shielded vaginal applicator, a shielded Fletcher Williamson applicator and an APBI balloon applicator. All applicators were modelled in Geant4 and MCNP6 using MG and CSG for dose calculations. CSG derived dose distributions were considered as reference and used to validate MG models by comparing dose distribution ratios. In general agreement within 1% for the dose calculations was observed for all applicators between MG and CSG and between codes when considering volumes inside the 25% isodose surface. When compared to CSG, MG required longer computation times by a factor of at least 2 for MC simulations using the same code. MCNP6 calculation times were more than ten times shorter than Geant4 in some cases. In conclusion we presented methods allowing for high fidelity modelling with results equivalent to CSG. To the best of our knowledge MG offers the most accurate representation of an irregular APBI balloon applicator.
Notes: [Fonseca, Gabriel Paiva; Yoriyaz, Helio] Inst Pesquisas Energet & Nucl IPEN CNEN SP, Sao Paulo, Brazil. [Fonseca, Gabriel Paiva; Landry, Guillaume; White, Shane; Reniers, Brigitte; Verhaegen, Frank] Maastricht Univ, Med Ctr, GROW Sch Oncol & Dev Biol, Dept Radiat Oncol MAASTRO, Maastricht, Netherlands. [Landry, Guillaume] Univ Munich, Fac Phys, Dept Med Phys, Munich, Germany. [D'Amours, Michel; Beaulieu, Luc] CHU Quebec, Dept Radiooncol, Quebec City, PQ, Canada. [D'Amours, Michel; Beaulieu, Luc] CHU Quebec, Ctr Rech, Quebec City, PQ, Canada. [D'Amours, Michel; Beaulieu, Luc] Univ Laval, Dept Phys Genie Phys & Opt, Quebec City, PQ, Canada. [D'Amours, Michel; Beaulieu, Luc] Univ Laval, Ctr Rech Cancerol, Quebec City, PQ, Canada. [Reniers, Brigitte] Hasselt Univ, CMK, Res Grp NuTeC, B-3590 Diepenbeek, Belgium. [Verhaegen, Frank] McGill Univ, Med Phys Unit, Montreal, PQ, Canada.
Keywords: mesh geometries; Monte Carlo; brachytherapy; CADMesh;mesh geometries; Monte Carlo; brachytherapy; CADMesh
Document URI: http://hdl.handle.net/1942/17753
ISSN: 0031-9155
e-ISSN: 1361-6560
DOI: 10.1088/0031-9155/59/19/5921
ISI #: 000342356800020
Rights: Content from this work may be used under the terms of the Creative Commons Attribution 3.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI. © 2014 Institute of Physics and Engineering in Medicine
Category: A1
Type: Journal Contribution
Validations: ecoom 2015
Appears in Collections:Research publications

Files in This Item:
File Description SizeFormat 
0031-9155_59_19_5921.pdf
  Restricted Access
2.96 MBAdobe PDFView/Open    Request a copy
Show full item record

SCOPUSTM   
Citations

4
checked on Sep 7, 2020

WEB OF SCIENCETM
Citations

6
checked on Apr 22, 2024

Page view(s)

102
checked on Apr 17, 2023

Download(s)

82
checked on Apr 17, 2023

Google ScholarTM

Check

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.