Title: BELdART-3: A national mailed audit program for Stereotaxy
Authors: YALVAC, Burak 
SCHROEYERS, Wouter 
REULENS, Nathalie 
SCHREURS, Sonja 
RENIERS, Brigitte 
Issue Date: 2017
Source: BHPA Symposium 2017 - Book of abstracts, p. 61-62
Abstract: BELdART-3, in continuation of BELdART-2, is now focusing on intracranial stereotactic radiosurgery (SRS) and extracranial Stereotactic Body Radiation Therapy (SBRT). The project is supported by the College of Medicine-radiotherapy and the FOD Healthcare. The radiation dose is measured using a combination of L-α-alanine-EMR dosimetry and Gafchromic® EBT3 films. The films were read using the triple channel dosimetry in the FilmQA Pro software developed by Ashland (Covington, Kentucky, USA) [1] and the “One-scan” procedure [2]. As for BELdART-2, we used alanine cylindrical pellets with height 2.8 mm and a diameter of 5 mm (Harwell Dosimeters LTD, Oxfordshire, UK). Some basic dosimetry is again done in water using the 3D printed support. Prior to the use of the anthropomorphic phantoms, the use of commercial alanine pellets was tested for small field dosimetry. Therefore, alanine pellets have been irradiated in combination with films in fields of 5x5 cm² to 1x1 cm² and the experiment has also been modelled in the EGSnrc Monte Carlo code. Once the limitations for the commercial alanine were known, they were included in the protocol for SRS delivery using the Stereotactic End-to-End Verification (STEEV) phantom (CIRS Inc, Norfolk, Virginia, USA) prepared with alanine and film. Furthermore, a custom support system for the phantom was developed to reduce positioning time and errors. The films were evaluated with gamma analysis [3] using the 5 %/1 mm tolerance levels with a lower dose threshold of 50 %. The correction factors for alanine in small fields were first estimated by Monte Carlo calculations, showing an over-response of the alanine > 2% for field sizes < 2 cm. This result was confirmed by measurements. Consequently, a target larger than 2 cm diameter is drawn for the combination film/alanine. The same sheet of EBT3 film is also used for a smaller target placed on another slice without alanine. The first tests show a ratio between the calculated dose and the alanine of 0.98 +/- 0.01. The film passing rate is 98 % for the large lesion in the STEEV phantom. The system of positioning has also improved both the accuracy and the speed of the positioning. The full procedure is still under testing. The combination of L-α-alanine-EMR dosimetry and film dosimetry show promising results for SRS audits. However, the field size must be larger than 2x2 cm² for accurate results of the alanine. The development of a similar protocol for SBRT is in progress. We would like to thank the physicists from the Jessa hospital and the Cliniques Universitaires Saint-Luc for their time helping us to develop the protocol and the Cancer Plan, FOD Healthcare and the College of Medicine for the financial and scientific support. In addition, we would like to thank the members from the steering committee for their support and scientific contribution. No conflict of interest.
Keywords: Beldart; SRS; SBRT; VMAT; dose verification; EPR dosimetry; film dosimetry
Document URI: http://hdl.handle.net/1942/26147
Category: C2
Type: Proceedings Paper
Appears in Collections:Research publications

Files in This Item:
File Description SizeFormat 
BHPA2017_Abstract_Burak_Yalvac_finaal.pdfPublished version340.26 kBAdobe PDFView/Open
Show full item record

Page view(s)

40
checked on Sep 6, 2022

Download(s)

8
checked on Sep 6, 2022

Google ScholarTM

Check


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.