Please use this identifier to cite or link to this item: http://hdl.handle.net/1942/31443
Title: Mechanical evaluation of the Bravos afterloader system for HDR brachytherapy
Authors: Bellezzo, Murillo
Baeza, Jose A.
Voncken, Robert
RENIERS, Brigitte 
Verhaegen, Frank
Fonseca, Gabriel P.
Issue Date: 2019
Publisher: ELSEVIER SCIENCE INC
Source: Brachytherapy, 18 (6) , p. 852 -862
Abstract: PURPOSE: The Bravos afterloader system was released by Varian Medical Systems in October of 2018 for high-dose-rate brachytherapy with Ir-192 sources, containing new features such as the CamScale (a new device for daily quality assurance and system recalibration), channel length verification, and different settings for rigid and flexible applicators. This study mechanically evaluated the Bravos system precision and accuracy for clinically relevant scenarios, using dummy sources. METHODS AND MATERIALS: The system was evaluated after three sets of experiments: (1) The CamScale was used to verify inter- and infra-channel dwelling variability and system calibration; (2) A high-speed camera was used to verify the source simulation cable movement inside a transparent quality assurance device, where dwell positions, dwell times, transit times, speed profiles, and accelerations were measured; (3) The source movement inside clinical applicators was captured with an imaging panel while being exposed to an external kV source. Measured and planned dwell positions and times were compared. RESULTS: Maximum deviations between planned and measured dwell positions and times for the source cable were 0.4 mm for the CamScale measurements and 0.07 seconds for the high-speed camera measurements. Mean dwell position deviations inside clinical applicators were below 1.2 mm for all applicators except the ring that required an offset correction of 1 mm to achieve a mean deviation of 0.4 mm. CONCLUSIONS: Features of the Bravos afterloader system provide a robust and precise treatment delivery. All measurements were within manufacturer specifications.
Notes: Fonseca, GP (reprint author), Maastricht Univ, Med Ctr, GROW Sch Oncol & Dev Biol, Dept Radiat Oncol MAASTRO, NL-6201 BN Maastricht, Netherlands.
gabriel.paivafonseca@maastro.nl
Keywords: brachytherapy;Afterloader;Commissioning
Document URI: http://hdl.handle.net/1942/31443
ISSN: 1538-4721
e-ISSN: 1873-1449
DOI: 10.1016/j.brachy.2019.06.005
ISI #: WOS:000500369700016
Rights: 2019 American Brachytherapy Society. Published by Elsevier Inc. All rights reserved.
Category: A1
Type: Journal Contribution
Validations: ecoom 2020
Appears in Collections:Research publications

Show full item record

SCOPUSTM   
Citations

1
checked on Sep 2, 2020

WEB OF SCIENCETM
Citations

9
checked on Apr 22, 2024

Page view(s)

28
checked on Sep 7, 2022

Google ScholarTM

Check

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.