Please use this identifier to cite or link to this item: http://hdl.handle.net/1942/39524
Title: Reliability, validity and clinical usability of a robotic assessment of finger proprioception in persons with multiple sclerosis
Authors: Zbytniewska-Mégret, Monika
Kanzler, Christoph M.
RAATS, Joke 
Gassert, Roger
YILMAZER, Cigdem 
FEYS, Peter 
Lambercy, Olivier
LAMERS, Ilse 
Issue Date: 2023
Publisher: 
Source: Multiple Sclerosis and Related Disorders, 70 (Art N° 104521)
Abstract: Background: Multiple sclerosis often leads to proprioceptive impairments of the hand. However, it is challenging to objectively assess such deficits using clinical methods, thereby also impeding accurate tracking of disease progression and hence the application of personalized rehabilitation approaches. Objective: We aimed to evaluate test-retest reliability, validity, and clinical usability of a novel robotic assessment of hand proprioceptive impairments in persons with multiple sclerosis (pwMS). Methods: The assessment was implemented in an existing one-degree of freedom end-effector robot (ETH MIKE) acting on the index finger metacarpophalangeal joint. It was performed by 45 pwMS and 59 neurologically intact controls. Additionally, clinical assessments of somatosensation, somatosensory evoked potentials and usability scores were collected in a subset of pwMS. Results: The test-retest reliability of robotic task metrics in pwMS was good (ICC=0.69-0.87). The task could identify individuals with impaired proprioception, as indicated by the significant difference between pwMS and controls, as well as a high impairment classification agreement with a clinical measure of proprioception (85.00-86.67%). Proprioceptive impairments were not correlated with other modalities of somatosensation. The usability of the assessment system was satisfactory (System Usability Scale ≥73.10). Conclusion: The proposed assessment is a promising alternative to commonly used clinical methods and will likely contribute to a better understanding of proprioceptive impairments in pwMS.
Keywords: Multiple sclerosis;Assessment;Rehabilitation;Somatosensory function;Proprioception
Document URI: http://hdl.handle.net/1942/39524
ISSN: 2211-0348
e-ISSN: 2211-0356
DOI: 10.1016/j.msard.2023.104521
ISI #: 000926869700001
Category: A1
Type: Journal Contribution
Appears in Collections:Research publications

Files in This Item:
File Description SizeFormat 
1-s2.0-S2211034823000251-main.pdfPublished version3.29 MBAdobe PDFView/Open
Show full item record

WEB OF SCIENCETM
Citations

3
checked on Apr 30, 2024

Google ScholarTM

Check

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.